Modular Spectrometers

Modular Spectroscopy Overview chart:

MONOCHROMATORS AND SPECTROGRAPHS


MODEL

FOCAL LENGTH (mm)

F/#

RESOLUTION (nm)

DISPERSION (nm/mm)

TURRET CONFIGURATION

SPECIAL FEATURES

SMALL SERIES

9030

100

F/3.2

1

8

Single Grating 32mm x 32mm

Compact Monochromator

9030DA

100x2

F/3.2

0.5

4

Single Grating 32mm x 32mm

Double Additive

9030DS

100x2

F/3.2

1

Effectively 0

Single Grating 32mm x 32mm

Double Subtractive

9072

125

F/3.5

0.4

7

Triple Grating 30mm x 30mm

Motorized Compact Monochromator

9072DA

125x2

F/3.5

0.2

3.5

Triple Grating 30mm x 30mm

Double Additive

9072DS

125x2

F/3.5

0.4

Effectively 0

Triple Grating 30mm x 30mm

Double Subtractive

STANDARD SERIES

9010

200

F/3.5

0.4

4

Double Grating 50mm x 50mm

Manual Monochromator/Spectrograph

9055

250

F/3.5

0.2

4

Triple Grating 50mm x 50mm

Motorized Monochromator/Spectrograph

9055F

250

F/2.5

0.4

3.3

Double Grating 64mm x 64mm

Fast Version Motorized Monochromator/Spectrograph

9055DX

250x2

F/3.5

0.1

1.7

Triple Grating 50mm x 50mm

Motorized Double Additive/Subtractive

9055DX-i

250x2

F/3.5

0.09

1.7

Triple Grating 50mm x 50mm

Motorized Imaging Monochromator/Spectrograph

9057

457

F/8

0.2

1.79

Triple Grating 50mm x 50mm

Half Meter Motorized Monochromator/Spectrograph

9057i

457

F/8

0.1

1.79

Triple Grating 50mm x 50mm

Half Meter Motorized Imaging Monochromator/Spectrograph

9057F

457

F/5.9

0.1

1.79

Triple Grating 64mm x 64mm

Fast Version Half Meter Motorized Monochromator/Spectrograph

HIGH RESOLUTION SERIES

9040

550

F/6.9

0.03

1.53

Up to 3 gratings 64mm x 64mm

Half Meter High Resolution

9040F

550

F/5.2

0.03

1.53

Single Grating 84mm x 84mm

Half Meter High Resolution Fast Version

9040EF

550

F/4.0

0.05

1.53

Single Grating 110mm x 110mm

Half Meter High Performace

9040-DS

550

F/6.9

0.03

Effectively 0

Single Grating 64mm x 64mm

Half Meter High Resolution Double Subtractive

9040DA

550

F/6.9

0.015

4

Single Grating 64mm x 64mm

Half Meter High Resolution Double Additive

9490

1000

F/13

0.017

0.83

Up to 3 gratings 64mm x 64mm

One Meter High Resolution

9490SD

1000

F/13

0.009

0.42

Up to 3 gratings 64mm x 64mm

One Meter High Resolution Single/Double Pass

9150

1500

F/12

0.013

0.5

Single Grating 110mm x 110mm

One and a Half Meter High Resolution

9150SD

1500x2

F/12

0.0063

0.25

Single Grating 110mm x 110mm

One and a Half Meter Single/Double Pass

9150X

1500

F/25

N/A

1pm/1px

Special gratings

One and a Half Meter Echelle Spectrometer

 


 

X

Product Category

View Filter

x
Clear all x

Filter Option



This is an upgrade to Sciencetech's High Resolution monochromators, including the 9040, 9490, and 9150 (for both the Single and Double Pass configurations of these systems). It moves the input slit from the default position on the front of the monochromator to the side or lateral position.
Please note that this upgrade does not allow both input ports to be used, it only moves the position. If you require both input ports and the flipping mirror to select between them, please see DPIN-HI products.
This is an upgrade to Sciencetech's High Resolution monochromators, including the 9040, 9490, and 9150 (for both the Single and Double Pass configurations of these systems). It moves the input slit from the default position on the front of the monochromator to the side or lateral position.
Please note that this upgrade does not allow both input ports to be used, it only moves the position. If you require both input ports and the flipping mirror to select between them, please see DPIN-HI products.
449.00 449.0 USD
Price Valid only in North America


This upgrade is for Sciencetech's High Resolution line of monochromators, including the 9040, 9490, and 9150 (for both the single and double pass options for those systems). It moves the output port from the front of the monochromator to the side or lateral position. This allows different orientations of the system in relation to various sample chambers or detectors, and may be required depending on the size of the detector you are using due to the close proximity and limited space between the input and output slit on the front plate.
Please note that this upgrade does not allow both output ports to be used; it only moves the default output from the front of the system to the side. If you require both output ports be usable, please see DPOUT-HI products.
This upgrade is for Sciencetech's High Resolution line of monochromators, including the 9040, 9490, and 9150 (for both the single and double pass options for those systems). It moves the output port from the front of the monochromator to the side or lateral position. This allows different orientations of the system in relation to various sample chambers or detectors, and may be required depending on the size of the detector you are using due to the close proximity and limited space between the input and output slit on the front plate.
Please note that this upgrade does not allow both output ports to be used; it only moves the default output from the front of the system to the side. If you require both output ports be usable, please see DPOUT-HI products.
364.00 364.0 USD
Price Valid only in North America


The 9030DA double additive monochromator consists of two 9030 monochromators placed in series. Stray light is drastically reduced in a double monochromator, and in the 9030DA it is further minimized by the use of concave holographic gratings. The inbound light is dispersed by the first grating and again by the second. Both gratings rotate simultaneously in the same direction, and the monochromators are coupled so that the dispersion is double that of a single 9030 monochromator. Bandpass is half of the corresponding 9030 monochromator for the same output slit, and resolution using a 100 µm slit is 0.4 nm.

Mirror coatings have been chosen for this model to enhance performance in the UV-VIS spectral region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a UV to VIS range to pair with the UV/VIS mirror coatings on this model. Two gratings are required.

The 9030DA is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00
In Stock
The 9030DA double additive monochromator consists of two 9030 monochromators placed in series. Stray light is drastically reduced in a double monochromator, and in the 9030DA it is further minimized by the use of concave holographic gratings. The inbound light is dispersed by the first grating and again by the second. Both gratings rotate simultaneously in the same direction, and the monochromators are coupled so that the dispersion is double that of a single 9030 monochromator. Bandpass is half of the corresponding 9030 monochromator for the same output slit, and resolution using a 100 µm slit is 0.4 nm.

Mirror coatings have been chosen for this model to enhance performance in the UV-VIS spectral region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a UV to VIS range to pair with the UV/VIS mirror coatings on this model. Two gratings are required.

The 9030DA is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00 9399.0 USD
Price Valid only in North America


The 9030DA double additive monochromator consists of two 9030 monochromators placed in series. Stray light is drastically reduced in a double monochromator, and in the 9030DA it is further minimized by the use of concave holographic gratings. The inbound light is dispersed by the first grating and again by the second. Both gratings rotate simultaneously in the same direction, and the monochromators are coupled so that the dispersion is double that of a single 9030 monochromator. Bandpass is half of the corresponding 9030 monochromator for the same output slit, and resolution using a 100 µm slit is 0.4 nm.

Mirror coatings have been chosen for this model to enhance performance in the VIS-NIR spectral region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a VIS to NIR range to pair with the VIS/IR mirror coatings on this model. Two gratings are required.

The 9030DA is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00
In Stock
The 9030DA double additive monochromator consists of two 9030 monochromators placed in series. Stray light is drastically reduced in a double monochromator, and in the 9030DA it is further minimized by the use of concave holographic gratings. The inbound light is dispersed by the first grating and again by the second. Both gratings rotate simultaneously in the same direction, and the monochromators are coupled so that the dispersion is double that of a single 9030 monochromator. Bandpass is half of the corresponding 9030 monochromator for the same output slit, and resolution using a 100 µm slit is 0.4 nm.

Mirror coatings have been chosen for this model to enhance performance in the VIS-NIR spectral region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a VIS to NIR range to pair with the VIS/IR mirror coatings on this model. Two gratings are required.

The 9030DA is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00 9399.0 USD
Price Valid only in North America


The 9030DS consists of two 100 mm scanning monochromators placed in series in subtractive mode. Bandpass is determined by the width of an intermediate slit, but the 9030DS has a net output dispersion of zero because the monochromators are optically coupled so that the dispersion of the second grating cancels that of the first. This leads to homogenized wavelengths at the output port - there is no spectrum, only a single undispersed slit image - which allows the unit to act as a tunable bandpass or notch filter.

Stray light of all sorts is drastically reduced in a double monochromator, and in the 9030DS it is further minimized by the use of concave holographic gratings. It is ideal for applications where stray light is a problem - where there may be a strong scattered background, such as in spectroradiometry, Raman, or measurements of solar radiation.

The 9030DS can be used as a tunable notch filter to remove one particular wavelength while allowing all others over the bandpass to pass through, making it an ideal instrument to pre-filter out the strong laser excitation wavelength in a fluorescence or Raman system so that the weaker output signals can be observed.

The 9030DS is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

The Model 9030DS can also be used to pass or filter out several selected wavelengths by using a custom intermediate multi-slit mask instead of a single width intermediate slit. Please contact a Sciencetech Applications Specialist for such custom wavelength filtering.

Mirror coatings have been chosen for this model to enhance performance in the UV-VIS region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a UV to VIS range to pair with the UV/VIS mirror coatings on this model. Two gratings are required.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00
In Stock
The 9030DS consists of two 100 mm scanning monochromators placed in series in subtractive mode. Bandpass is determined by the width of an intermediate slit, but the 9030DS has a net output dispersion of zero because the monochromators are optically coupled so that the dispersion of the second grating cancels that of the first. This leads to homogenized wavelengths at the output port - there is no spectrum, only a single undispersed slit image - which allows the unit to act as a tunable bandpass or notch filter.

Stray light of all sorts is drastically reduced in a double monochromator, and in the 9030DS it is further minimized by the use of concave holographic gratings. It is ideal for applications where stray light is a problem - where there may be a strong scattered background, such as in spectroradiometry, Raman, or measurements of solar radiation.

The 9030DS can be used as a tunable notch filter to remove one particular wavelength while allowing all others over the bandpass to pass through, making it an ideal instrument to pre-filter out the strong laser excitation wavelength in a fluorescence or Raman system so that the weaker output signals can be observed.

The 9030DS is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

The Model 9030DS can also be used to pass or filter out several selected wavelengths by using a custom intermediate multi-slit mask instead of a single width intermediate slit. Please contact a Sciencetech Applications Specialist for such custom wavelength filtering.

Mirror coatings have been chosen for this model to enhance performance in the UV-VIS region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a UV to VIS range to pair with the UV/VIS mirror coatings on this model. Two gratings are required.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00 9399.0 USD
Price Valid only in North America


The 9030DS consists of two 100 mm scanning monochromators placed in series in subtractive mode. Bandpass is determined by the width of an intermediate slit, but the 9030DS has a net output dispersion of zero because the monochromators are optically coupled so that the dispersion of the second grating cancels that of the first. This leads to homogenized wavelengths at the output port - there is no spectrum, only a single undispersed slit image - which allows the unit to act as a tunable bandpass or notch filter.

Stray light of all sorts is drastically reduced in a double monochromator, and in the 9030DS it is further minimized by the use of concave holographic gratings. It is ideal for applications where stray light is a problem - where there may be a strong scattered background, such as in spectroradiometry, Raman, or measurements of solar radiation.

The 9030DS can be used as a tunable notch filter to remove one particular wavelength while allowing all others over the bandpass to pass through, making it an ideal instrument to pre-filter out the strong laser excitation wavelength in a fluorescence or Raman system so that the weaker output signals can be observed.

The 9030DS is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

The Model 9030DS can also be used to pass or filter out several selected wavelengths by using a custom intermediate multi-slit mask instead of a single width intermediate slit. Please contact a Sciencetech Applications Specialist for such custom wavelength filtering.

Mirror coatings have been chosen for this model to enhance performance in the VIS-NIR region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a VIS to NIR range to pair with the VIS/IR mirror coatings on this model. Two gratings are required.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00
In Stock
The 9030DS consists of two 100 mm scanning monochromators placed in series in subtractive mode. Bandpass is determined by the width of an intermediate slit, but the 9030DS has a net output dispersion of zero because the monochromators are optically coupled so that the dispersion of the second grating cancels that of the first. This leads to homogenized wavelengths at the output port - there is no spectrum, only a single undispersed slit image - which allows the unit to act as a tunable bandpass or notch filter.

Stray light of all sorts is drastically reduced in a double monochromator, and in the 9030DS it is further minimized by the use of concave holographic gratings. It is ideal for applications where stray light is a problem - where there may be a strong scattered background, such as in spectroradiometry, Raman, or measurements of solar radiation.

The 9030DS can be used as a tunable notch filter to remove one particular wavelength while allowing all others over the bandpass to pass through, making it an ideal instrument to pre-filter out the strong laser excitation wavelength in a fluorescence or Raman system so that the weaker output signals can be observed.

The 9030DS is a fully integrated unit with the two diffraction gratings coupled mechanically for accurate positioning, adjusted manually using a thumb-wheel. It can be upgraded to be motorized and controlled by a computer with the 9030 Motorized Upgrade. Only one motor and drive unit is required for both monochromators, enabling control via Sciencetech Sci-Mono Active-X library or full featured Sci-Spec software.

The Model 9030DS can also be used to pass or filter out several selected wavelengths by using a custom intermediate multi-slit mask instead of a single width intermediate slit. Please contact a Sciencetech Applications Specialist for such custom wavelength filtering.

Mirror coatings have been chosen for this model to enhance performance in the VIS-NIR region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a VIS to NIR range to pair with the VIS/IR mirror coatings on this model. Two gratings are required.

Please add a minimum of three slits (for the input, output, and intermediate position) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
9,399.00 9399.0 USD
Price Valid only in North America


The Sciencetech 9030 monochromator is a compact, single grating, scanning monochromator with 100 mm focal length (both output and input). It has a manual wavelength selector dial with an optional external motor drive for computerized control. The compact design is made possible by its Seya-Namioka optical layout, using a single holographic concave grating and f/3.2 aperture. The input and output ports are purposely placed optically in-line, making it a simple addition to a straight line beam in an optical system. Being a scanning monochromator, it does not have a flat output field and hence cannot be operated as a spectrograph with a multi-channel detector mounted at the output port. However, it can be used as an excitation wavelength tunable light source.

Mirror coatings have been chosen for this model to enhance performance in the UV-VIS region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a UV to VIS range to pair with the UV/VIS mirror coatings on this model.

Please add a minimum of two slits (for the input and output) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
3,639.00
The Sciencetech 9030 monochromator is a compact, single grating, scanning monochromator with 100 mm focal length (both output and input). It has a manual wavelength selector dial with an optional external motor drive for computerized control. The compact design is made possible by its Seya-Namioka optical layout, using a single holographic concave grating and f/3.2 aperture. The input and output ports are purposely placed optically in-line, making it a simple addition to a straight line beam in an optical system. Being a scanning monochromator, it does not have a flat output field and hence cannot be operated as a spectrograph with a multi-channel detector mounted at the output port. However, it can be used as an excitation wavelength tunable light source.

Mirror coatings have been chosen for this model to enhance performance in the UV-VIS region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a UV to VIS range to pair with the UV/VIS mirror coatings on this model.

Please add a minimum of two slits (for the input and output) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
3,639.00 3639.0 USD
Price Valid only in North America


The Sciencetech 9030 monochromator is a compact, single grating, scanning monochromator with 100 mm focal length (both output and input). It has a manual wavelength selector dial with an optional external motor drive for computerized control. The compact design is made possible by its Seya-Namioka optical layout, using a single holographic concave grating and f/3.2 aperture. The input and output ports are purposely placed optically in-line, making it a simple addition to a straight line beam in an optical system. Being a scanning monochromator, it does not have a flat output field and hence cannot be operated as a spectrograph with a multi-channel detector mounted at the output port. However, it can be used as an excitation wavelength tunable light source.

Mirror coatings have been chosen for this model to enhance performance in the VIS-IR region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a VIS to NIR range to pair with the VIS/IR mirror coatings on this model.

Please add a minimum of two slits (for the input and output) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
3,539.00
The Sciencetech 9030 monochromator is a compact, single grating, scanning monochromator with 100 mm focal length (both output and input). It has a manual wavelength selector dial with an optional external motor drive for computerized control. The compact design is made possible by its Seya-Namioka optical layout, using a single holographic concave grating and f/3.2 aperture. The input and output ports are purposely placed optically in-line, making it a simple addition to a straight line beam in an optical system. Being a scanning monochromator, it does not have a flat output field and hence cannot be operated as a spectrograph with a multi-channel detector mounted at the output port. However, it can be used as an excitation wavelength tunable light source.

Mirror coatings have been chosen for this model to enhance performance in the VIS-IR region. There are 5 grating choices (sold separately) covering from UV to near-IR; choose a VIS to NIR range to pair with the VIS/IR mirror coatings on this model.

Please add a minimum of two slits (for the input and output) separately. Several removable fixed-width slit sizes are available to choose from ranging from 50 μm to 3 mm width. Please note that a 50μm slit may introduce some artifacts when used with the 9030 - this model is recommended to be used with slits of 100μm or larger for best performance in most applications. Additional removable slits and custom width slits can also be purchased.
Available slits include:
Model Slit Width
SF005 - 50 μm
SF010 - 100μm
SF025 - 250μm
SF050 - 500μm
SF100 - 1mm
SF200 - 2mm
SF300 - 3mm
3,539.00 3539.0 USD
Price Valid only in North America


The 9040 is Sciencetech's 0.5 m focal length monochromator. This model of monochromator accommodates a grating on a removable kinematic mount, allowing the user to manually swap one grating for another without losing alignment. This version can accommodate 64 x 64 mm, 84 x 84 mm, and 102 x 102 mm gratings. Please note that this product includes one slit, but does not include any additional slits or a grating, these must be purchased separately. Please speak with a technical sales member for assistance in selecting the right components for your system. SciMono is included. For Sciencetech's full feature spectroscopy software, please see SciSpec (not included).

Please see the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
The 9040 is Sciencetech's 0.5 m focal length monochromator. This model of monochromator accommodates a grating on a removable kinematic mount, allowing the user to manually swap one grating for another without losing alignment. This version can accommodate 64 x 64 mm, 84 x 84 mm, and 102 x 102 mm gratings. Please note that this product includes one slit, but does not include any additional slits or a grating, these must be purchased separately. Please speak with a technical sales member for assistance in selecting the right components for your system. SciMono is included. For Sciencetech's full feature spectroscopy software, please see SciSpec (not included).

Please see the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
31,299.00 31299.0 USD
Price Valid only in North America


The 9040 is Sciencetech's 0.5 m focal length monochromator. This model of monochromator accommodates up to 3 gratings on a rotational turret to automatically scan over all gratings through software. This version can accommodate 64 x 64 mm, 84 x 84 mm, and 102 x 102 mm gratings. Please note that this product includes one slit, but does not include any additional slits or a grating, these must be purchased separately. Please speak with a technical sales member for assistance in selecting the right components for your system. SciMono is included. For Sciencetech's full feature spectroscopy software, please see SciSpec (not included).

Please see the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
The 9040 is Sciencetech's 0.5 m focal length monochromator. This model of monochromator accommodates up to 3 gratings on a rotational turret to automatically scan over all gratings through software. This version can accommodate 64 x 64 mm, 84 x 84 mm, and 102 x 102 mm gratings. Please note that this product includes one slit, but does not include any additional slits or a grating, these must be purchased separately. Please speak with a technical sales member for assistance in selecting the right components for your system. SciMono is included. For Sciencetech's full feature spectroscopy software, please see SciSpec (not included).

Please see the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
69,999.00 69999.0 USD
Price Valid only in North America


The 9040DX is a Double Additive or Subtractive design of the half meter 9040. Special care has been taken to design the system to minimize and eliminate stray light whenever possible.
A double subtractive monochromator operates to create a monochromatic light source by the 2nd monochromator cancelling the dispersion caused by the 1st monochromator. This allows a Double Subtractive system to operate as a tunable notch filter for Raman applications. The double additive is used for improved resolution. Please see the technical datasheet or contact a Sciencetech Technical Representative for additional information.

Please note that 2 gratings are required for this system, minimum, but are not included. This version can accommodate 64 x 64 mm, 84 x 84 mm, and 102 x 102 mm gratings. Please note that this product includes one slit, but does not include any additional slits or a grating, these must be purchased separately. SciMono is included. For Sciencetech's full feature spectroscopy software, please see SciSpec (not included).

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
The 9040DX is a Double Additive or Subtractive design of the half meter 9040. Special care has been taken to design the system to minimize and eliminate stray light whenever possible.
A double subtractive monochromator operates to create a monochromatic light source by the 2nd monochromator cancelling the dispersion caused by the 1st monochromator. This allows a Double Subtractive system to operate as a tunable notch filter for Raman applications. The double additive is used for improved resolution. Please see the technical datasheet or contact a Sciencetech Technical Representative for additional information.

Please note that 2 gratings are required for this system, minimum, but are not included. This version can accommodate 64 x 64 mm, 84 x 84 mm, and 102 x 102 mm gratings. Please note that this product includes one slit, but does not include any additional slits or a grating, these must be purchased separately. SciMono is included. For Sciencetech's full feature spectroscopy software, please see SciSpec (not included).

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
58,599.00 58599.0 USD
Price Valid only in North America


The Sciencetech 9055 is a 250mm asymmetric focal length modified Czerny-Turner monochromator. It has a flat field output for use as a spectrograph. This model has an f/3.5 aperture (for the "Fast" version with f/2.5, please see the 9055F). It has a computer-controlled interchangeable turret for automatic grating and wavelength selection.
Both the input and output ports are located on the front faceplate of the monochromator, and the system includes one SS80 slit at the input port. CE compliant.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
8,453.00
In Stock
The Sciencetech 9055 is a 250mm asymmetric focal length modified Czerny-Turner monochromator. It has a flat field output for use as a spectrograph. This model has an f/3.5 aperture (for the "Fast" version with f/2.5, please see the 9055F). It has a computer-controlled interchangeable turret for automatic grating and wavelength selection.
Both the input and output ports are located on the front faceplate of the monochromator, and the system includes one SS80 slit at the input port. CE compliant.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
8,453.00 8453.0 USD
Price Valid only in North America


9055DX is a motorized double 0.25 m monochromator that can be configured as additive or subtractive. Double monochromators have the advantage of extremely low stray light levels by virtue of using two gratings. An intermediate slit between the two units, used to define the wavelengths that pass to the second unit, also masks off light scattered within the body of the first unit. In double additive operation the 9055DX effective focal length (0.5 m), dispersion, and resolution (1.7 nm/mm and 0.09 nm with a 1200 l/mm grating) are doubled compared to a single unit, making this mode useful where high resolution is important. In double subtractive operation there is no dispersion in the output beam and only a single slit image appears at the output instead of a broad spread of spectral images. This allows the system to operate as a tunable bandpass or notch filter.

This system includes two manual slits (one input and one intermediate). Motorized upgrades and additional slits may be purchased separately. Gratings also need to be purchased separately, with a minimum of 2 gratings required for the system to operate. Please speak with your Sciencetech technical sales member to ensure proper grating selection.

Please see the technical datasheet for additional information.

This system includes 2 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
22,999.00
In Stock
9055DX is a motorized double 0.25 m monochromator that can be configured as additive or subtractive. Double monochromators have the advantage of extremely low stray light levels by virtue of using two gratings. An intermediate slit between the two units, used to define the wavelengths that pass to the second unit, also masks off light scattered within the body of the first unit. In double additive operation the 9055DX effective focal length (0.5 m), dispersion, and resolution (1.7 nm/mm and 0.09 nm with a 1200 l/mm grating) are doubled compared to a single unit, making this mode useful where high resolution is important. In double subtractive operation there is no dispersion in the output beam and only a single slit image appears at the output instead of a broad spread of spectral images. This allows the system to operate as a tunable bandpass or notch filter.

This system includes two manual slits (one input and one intermediate). Motorized upgrades and additional slits may be purchased separately. Gratings also need to be purchased separately, with a minimum of 2 gratings required for the system to operate. Please speak with your Sciencetech technical sales member to ensure proper grating selection.

Please see the technical datasheet for additional information.

This system includes 2 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
22,999.00 22999.0 USD
Price Valid only in North America


9055DXi is a motorized double 0.25 m monochromator that can be configured as additive or subtractive. The imaging option 9055DXi uses toroidal optics to compensate for astigmatism inherent in Czerney-Turner type monochromators and spectrographs. Double monochromators have the advantage of extremely low stray light levels by virtue of using two gratings. An intermediate slit between the two units, used to define the wavelengths that pass to the second unit, also masks off light scattered within the body of the first unit. In double additive operation the 9055DX effective focal length (0.5 m), dispersion, and resolution (1.7 nm/mm and 0.09 nm with a 1200 l/mm grating) are doubled compared to a single unit, making this mode useful where high resolution is important. In double subtractive operation there is no dispersion in the output beam and only a single slit image appears at the output instead of a broad spread of spectral images. This allows the system to operate as a tunable bandpass or notch filter.

This system includes two manual slits (one input and one intermediate). Motorized upgrades and additional slits may be purchased separately. Gratings also need to be purchased separately, with a minimum of 2 gratings required for the system to operate. Please speak with your Sciencetech technical sales member to ensure proper grating selection.

Please see the technical datasheet for additional information.

This system includes 2 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
23,180.00
In Stock
9055DXi is a motorized double 0.25 m monochromator that can be configured as additive or subtractive. The imaging option 9055DXi uses toroidal optics to compensate for astigmatism inherent in Czerney-Turner type monochromators and spectrographs. Double monochromators have the advantage of extremely low stray light levels by virtue of using two gratings. An intermediate slit between the two units, used to define the wavelengths that pass to the second unit, also masks off light scattered within the body of the first unit. In double additive operation the 9055DX effective focal length (0.5 m), dispersion, and resolution (1.7 nm/mm and 0.09 nm with a 1200 l/mm grating) are doubled compared to a single unit, making this mode useful where high resolution is important. In double subtractive operation there is no dispersion in the output beam and only a single slit image appears at the output instead of a broad spread of spectral images. This allows the system to operate as a tunable bandpass or notch filter.

This system includes two manual slits (one input and one intermediate). Motorized upgrades and additional slits may be purchased separately. Gratings also need to be purchased separately, with a minimum of 2 gratings required for the system to operate. Please speak with your Sciencetech technical sales member to ensure proper grating selection.

Please see the technical datasheet for additional information.

This system includes 2 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
23,180.00 23180.0 USD
Price Valid only in North America


The Sciencetech 9055F is a 250 mm focal length modified Czerny-Turner monochromator, with a flat field output for spectrograph use. This is designed as a faster version of the 9055 with larger optics for higher light throughput and f/2.5 aperture. All 9055Fs have a computer controlled, motorized turret for automatic grating and wavelength selection. Both input and output ports are situated on the front faceplate of the monochromator/spectrograph, and the system includes one adjustable SS80 slit at the input port. The 9055F supports up to two 64 x 64 mm (clear aperture) gratings.

Please see technical datasheet for further information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
11,339.00
In Stock
The Sciencetech 9055F is a 250 mm focal length modified Czerny-Turner monochromator, with a flat field output for spectrograph use. This is designed as a faster version of the 9055 with larger optics for higher light throughput and f/2.5 aperture. All 9055Fs have a computer controlled, motorized turret for automatic grating and wavelength selection. Both input and output ports are situated on the front faceplate of the monochromator/spectrograph, and the system includes one adjustable SS80 slit at the input port. The 9055F supports up to two 64 x 64 mm (clear aperture) gratings.

Please see technical datasheet for further information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
11,339.00 11339.0 USD
Price Valid only in North America


The Sciencetech 9055i is a 250 mm asymmetric focal length modified Czerny-Turner monochromator. It has a flat field output for use as a spectrograph, with an area of 25 mm wide and 12 mm high. This model has an f/3.5 aperture. It has a computer-controlled interchangeable turret for automatic grating and wavelength selection. Both the input and output ports are located on the front faceplate of the monochromator, and the system includes one SS80 slit at the input port. CE compliant.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
10,020.00
In Stock
The Sciencetech 9055i is a 250 mm asymmetric focal length modified Czerny-Turner monochromator. It has a flat field output for use as a spectrograph, with an area of 25 mm wide and 12 mm high. This model has an f/3.5 aperture. It has a computer-controlled interchangeable turret for automatic grating and wavelength selection. Both the input and output ports are located on the front faceplate of the monochromator, and the system includes one SS80 slit at the input port. CE compliant.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
10,020.00 10020.0 USD
Price Valid only in North America


The 9057 is a 0.5 m monochromator and spectrograph with an f/8 aperture. This system is a Czerny-Turner design optimized to reduce aberrations.
This monochromator supports up to three 50x50mm gratings on a computer controlled turret. It includes a single input slit by default; please select appropriate output mounting (slit, detector or CCD mounting, etc...) when ordering. A fast version of the 9057, the 9057F, operates with 64x64mm gratings and offers better throughput with the same optical path length. Please discuss with your Sciencetech technical sales staff member to find if the 9057F is a better choice for your particular application.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
10,999.00
In Stock
The 9057 is a 0.5 m monochromator and spectrograph with an f/8 aperture. This system is a Czerny-Turner design optimized to reduce aberrations.
This monochromator supports up to three 50x50mm gratings on a computer controlled turret. It includes a single input slit by default; please select appropriate output mounting (slit, detector or CCD mounting, etc...) when ordering. A fast version of the 9057, the 9057F, operates with 64x64mm gratings and offers better throughput with the same optical path length. Please discuss with your Sciencetech technical sales staff member to find if the 9057F is a better choice for your particular application.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
10,999.00 10999.0 USD
Price Valid only in North America


The Sciencetech 9057F is a fast 0.5 m motorized Czerny-Turner monochromator (and spectrograph) with an f/5.9 aperture. This is a "faster" model with larger optics for higher light throughput compared to the standard 9057 with f/8 aperture. Both models utilize a computer controlled motorized turret for automatic grating and wavelength selection. The regular 9057 model supports up to three 50 x 50mm gratings whereas the larger optics 9057F model supports up to two 64 x 64mm (clear aperture) gratings.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).

11,499.00
In Stock
The Sciencetech 9057F is a fast 0.5 m motorized Czerny-Turner monochromator (and spectrograph) with an f/5.9 aperture. This is a "faster" model with larger optics for higher light throughput compared to the standard 9057 with f/8 aperture. Both models utilize a computer controlled motorized turret for automatic grating and wavelength selection. The regular 9057 model supports up to three 50 x 50mm gratings whereas the larger optics 9057F model supports up to two 64 x 64mm (clear aperture) gratings.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).

11,499.00 11499.0 USD
Price Valid only in North America


An imaging 0.5 m monochromator/spectrograph. It has an f/8 aperture and a 2D flat field output for multi-channel use. It has a nominal resolution of 0.1 nm. Special toroidal mirror makes additional astigmatism correction. This spectrograph provides superior image quality. The unique layout with toroidal mirrors is optimized for minimum coma while allowing high throughput and minimum stray light and vignetting. The 9057i model supports up to three 50 x 50 mm gratings.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
11,999.00
In Stock
An imaging 0.5 m monochromator/spectrograph. It has an f/8 aperture and a 2D flat field output for multi-channel use. It has a nominal resolution of 0.1 nm. Special toroidal mirror makes additional astigmatism correction. This spectrograph provides superior image quality. The unique layout with toroidal mirrors is optimized for minimum coma while allowing high throughput and minimum stray light and vignetting. The 9057i model supports up to three 50 x 50 mm gratings.

It Includes a Sci-Mono, a Windows based software module with a simple graphical user interface for controlling Sciencetech stepping monochromator wavelength selection. SciMono is provided free of charge with a Sciencetech motorized monochromator when SciSpec software has not been purchased. Sciencetech Software (Sci-Spec) is available at an additional cost. It is a standalone data acquisition and allows controlling the monochromator by itself. This system does not include gratings or a host computer. If you require any of these components, please speak with your technical sales staff member.

Refer to the technical datasheet for additional information.

This system includes 1 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
11,999.00 11999.0 USD
Price Valid only in North America


In the 9072DA, the inbound light is dispersed by the first grating then passed through a fixed intermediate slit where stray light is rejected and then spectrally dispersed again by the second grating.

The double dispersion of the double additive monochromators results in a doubling of the dispersion of a single monochromator, leading to better optical resolution. The bandpass is half that of a single 9072 monochromator.

One SS-80 slit is included, which can be attached to either the input or output port. A second slit can be added to the order if desired.
The fixed slit is not included in the 9072DA and can be selected from any of Sciencetech's SF-series of fixed slits and added to the order.
Gratings and grating holders are not included. The 9072DA accommodates 30 × 30 mm gratings, with up to three gratings per turret. Two of each selected grating must be purchased to fill both monochromators. One grating holder is required for each grating.

This system includes 2 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
19,995.00
In Stock
In the 9072DA, the inbound light is dispersed by the first grating then passed through a fixed intermediate slit where stray light is rejected and then spectrally dispersed again by the second grating.

The double dispersion of the double additive monochromators results in a doubling of the dispersion of a single monochromator, leading to better optical resolution. The bandpass is half that of a single 9072 monochromator.

One SS-80 slit is included, which can be attached to either the input or output port. A second slit can be added to the order if desired.
The fixed slit is not included in the 9072DA and can be selected from any of Sciencetech's SF-series of fixed slits and added to the order.
Gratings and grating holders are not included. The 9072DA accommodates 30 × 30 mm gratings, with up to three gratings per turret. Two of each selected grating must be purchased to fill both monochromators. One grating holder is required for each grating.

This system includes 2 IEC 60320 C13 compatible power cables. Region-specific version must be selected at the time of placing an order (see product 491-9001).
19,995.00 19995.0 USD
Price Valid only in North America
Copyright © Sciencetech